The effect of implant characteristics on the implant stability of immediately loaded single implant cases: A prospective study


1 Prosthetic Dentistry Unit, Federal Medical Centre, Abeokuta, Ogun State, Nigeria
2 Department of Restorative Dentistry, Lagos University Teaching Hospital and College of Medicine, University of Lagos, Lagos, Nigeria
3 State House Medical Centre, Aso Rock, Asokoro, Abuja-Fct, Nigeria
4 Department of Oral and Maxillofacial Surgery, Lagos University Teaching Hospital and College of Medicine, University of Lagos, Lagos, Nigeria

Date of Web Publication17-Dec-2018

Correspondence Address:
Dr. Kolawole O Obagbemiro
Prosthetic Dentistry Unit, Federal Medical Centre, Abeokuta, Ogun State
Nigeria
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jdi.jdi_12_18

Rights and Permissions
   Abstract 

Introduction: Factors that affect primary and ultimate implant stability include characteristics of the type of dental implant used, bone quality at the site of implant placement, insertion torque, as well as micromotions at the bone–implant interface. This study, however, sought out to determine the effect and relationship between relevant implant characteristics and implant stability in immediately loaded single implant cases using the Periotest® M handheld device.
Aim: To determine the effect of implant characteristics on the ultimate implant stability in immediately loaded single implant cases.
Results: At placement, 33 (94%) of implants had periotest values between −0.8 and 0 at placement. There is a general decrease in the number of patients with periotest values −0.8–0 and those with +1–9 from the initial placement to 6 months after placement. It was further observed that there is a negative correlation between the implant length, diameter, and the mean periotest values although this was not statistically significant.
Conclusion: There is a directly proportional relationship between implant characteristics and implant stability of immediately loaded implants.

Keywords: Implant characteristics, implant stability, immediate loading, prospective study


How to cite this article:
Obagbemiro KO, Ajayi YO, Akeredolu PA, Adeoye JA, Arotiba GT. The effect of implant characteristics on the implant stability of immediately loaded single implant cases: A prospective study. J Dent Implant 2018;8:48-53

How to cite this URL:
Obagbemiro KO, Ajayi YO, Akeredolu PA, Adeoye JA, Arotiba GT. The effect of implant characteristics on the implant stability of immediately loaded single implant cases: A prospective study. J Dent Implant [serial online] 2018 [cited 2019 May 21];8:48-53. Available from: http://www.jdionline.org/text.asp?2018/8/2/48/247580


   Introduction Top


While implant therapy has experienced numerous advances in implant designs as well as the surgical and restorative techniques, many clinicians have questioned whether or not the nonloaded healing period is still a valid prerequisite for success.[1] Although not applicable to all patients, the concept of immediately loaded dental implant is not novel but extends back to the 1960s, when implant dentistry was still in its relative infancy.[2] Its protocol advocates the loading of the prosthetic superstructure right after the implant has been placed in the bone.[1] Types of immediately loaded dental implant systems can either be of direct loading type or be of early functional loading type.[3],[4] In the direct loading type, the superstructure is attached to the implant within 24 h of implant placement, while in the early functional loading type, it is placed within days or weeks of implant placement. A fundamental requirement for immediate occlusal loading stability is adequate primary implant stability. Primary implant stability seems to be the most important determining factor on immediate implant loading as an immobile, functionally loaded implant is an essential ingredient to achieve osseointegration.[5] While stability was traditionally achieved through a period of undisturbed healing in conventional methods of implant placement, primary stability is now achieved via a mechanical phenomenon of screw stability and splinting.[6],[7],[8] Factors that affect primary and ultimate implant stability include characteristics of the type of dental implant used including implant length, diameter, and design, bone quality at the site of implant placement (monocortical/bicortical), insertion torque, as well as micromotions at the bone–implant interface. This study, however, sought out to determine the effect and relationship between relevant implant characteristics and implant stability in immediately loaded single implant cases.


   Materials and Methods Top


This is a longitudinal, nonrandomized clinical study implemented with a sample size of 35 subjects seeking to undergo single implant rehabilitation for missing teeth at a teaching hospital in Lagos, Nigeria. Recruitment of these individuals occurred consecutively within 1 year as they presented to the clinic with only those who met the inclusion criteria involved in the study.

Inclusion criteria

  • Patients with short edentulous span in the maxillary or mandibular arches with not more than two missing teeth
  • Patients aged 18–60 years
  • Patients edentulous for 4 months and above
  • Patients with no contraindications related to surgical or prosthetic procedures
  • Patients who gave their informed consent
  • Patients with normal-to-dense bone quality in the planned implant site determined by clinical inspection, palpation, and periapical radiograph
  • Patients with adequate bone volume to support an implant without the need of bone augmentation
  • Patient with edentulous site that permits the placement of an implant at least 10 mm in length.


Exclusion criteria

  • Patients with active infection in planned implant sites
  • Patients with systemic diseases such as uncontrolled diabetes, poorly controlled hypertension, bleeding disorders, and severely compromised immune system
  • Patients with a history of bruxism
  • Patients with a need for bone augmentation at the intended site
  • Patients with pregnancy or mentally unstable
  • Patients with poor oral hygiene, smoking, and not ready to quit the habit
  • Patients who cannot keep recall appointments
  • Patients who are poorly motivated in maintaining good oral hygiene
  • Grossly supraerupted opposing tooth that cannot allow implant prosthesis fabrication
  • Patients requiring orthodontic tooth movement
  • Second and third molars.


Following clinical and radiographic assessment of subjects who met the inclusion criteria, implants were placed by a single implantologist under adequate presurgical and surgical conditions. Implants of length 10, 11, and 13 mm and diameter 3.7, 4.1, and 4.7 mm were used. To ensure primary implant stability, implant screws were tightened with a manual torque wrench to ensure a torque of 40 N-cm for all participants [Figure 1]. Primary implant stability was also assessed using Periotest® M handheld device (Medizintechnik Gulden, Germany) [Figure 2] before loading.
Figure 1: Manual torque wrench

Click here to view
Figure 2: Periotest M handheld device

Click here to view


Subjects were recalled for subsequent visits and Periotest® M handheld electronic device was used to check implant stability at 1, 3, 6, and 9 months and 1 year. According to the manufacturer, interpretations of the periotest readings are as shown in [Table 1].[9]
Table 1: Periotest (R) value ranges interpretations

Click here to view



   Results Top


A total of 35 subjects were involved in this study comprising 23 females (66.0%) and 12 males (34%) who were within the age range of 21–60 years with a mean age of 43.14 ± 13.08 and median of 45.00 [Table 2]. Majority of subjects were, however, between ages 51 and 60 years. The mean torque of implants used is 45 N-cm. More implants were placed in the maxilla (71.4%) than in the mandible (28.6%). Most of the implants were 10 mm long (71.4%) and 4.1 mm wide (48.6%). [Table 3] summarizes the distribution of implant length and diameter of the immediately loaded implants. As regards implant stability, at placement, 33 (94%) participants had periotest values between −0.8 and 0 at placement. There is a general decrease in the number of patients with periotest values −0.8–0 and those with +1–9 from the initial placement to 6 months after placement. However, the number of patients with +10 and above increased from zero at the initial placement to 3 after 6 months. The frequency distribution of periotest values is summarized in [Table 4].
Table 2: Sociodemographic characteristics of participants

Click here to view
Table 3: Distribution of implant diameter by length of the 35 immediately loaded implants

Click here to view
Table 4: Frequency of periotest values at placement of implants to 1 year

Click here to view


Relating implant length to implant stability, 11.5 mm long implants had the highest mean periotest value after placement; however, after 1-year follow-up, the longer implants (13 mm) had the highest mean periotest value. No obvious trend was seen in the periotest value for each implant length over the period of follow-up although the lowest mean periotest values were obtained at 3 months [Table 5].
Table 5: Distribution of implant length by the mean periotest values from the time of placement to 1 year

Click here to view


There was a negative correlation between implant length and the mean periotest value with mean periotest score decreasing and implant stability increasing as implant length increasing although this was not statistically significant [Table 6].
Table 6: Correlation of implant length by total mean periotest score

Click here to view


As regards the relationship between implant diameter and implant stability, the mean periotest values at placement decrease as the diameter of the implant increases. However, at 1-year follow-up, the smallest diameter implant (3.7 mm) had the lowest mean periotest value [Table 7]. There was a negative correlation between implant diameter and the mean periotest score; however, this was not statistically significant (P < 0.05) [Table 8].
Table 7: Distribution of implant diameter by the mean periotest values from the time of placement to 1 year

Click here to view
Table 8: Correlation of implant diameter by total mean periotest score

Click here to view


There was a significant difference (P < 0.05) between the mean periotest score of implants placed in the mandible and the implants placed in the maxilla, with mandibular implant showing better stability [Table 9]. However, there was no statistically significant difference between the mean periotest scores of implants placed in the anterior and posterior segments of both jaws (P < 0.05) [Table 10].
Table 9: Distribution of site of implant placement (maxilla and mandible) by mean periotest score

Click here to view
Table 10: Distribution of location of implant placement (anterior and posterior) by mean periotest score

Click here to view


[Table 11] shows that 28 implants survived. Initially, 33 implants had an initial periotest value of − 0.8–0 (Category I), of which 28 survived; however, the two implants with an initial periotest value of +1–+9 failed. There was a statistical significant association between the periotest values at placement and implant survival with those who survived having a lower initial periotest value [Figure 3].
Table 11: Comparison of periotest values at placement and survival of implant

Click here to view
Figure 3: Relationship between periotest value and implant survival

Click here to view



   Discussion Top


This present study sought out to determine the stability of dental implants using the immediate loading protocol with a view to justifying its use in rehabilitating individuals with missing teeth. Implant length and diameter are two major implant-related factors determining the primary implant stability as well as the ultimate success of placement. The implant lengths used in this study are between 10 and 13 mm. Periotest values were used as a measure of implant stability in this study with negative and smaller periotest values indicating greater implant stability and higher values above +9, indicating reduction in implant stability and higher clinical degree of tooth mobility.[9] We found that 11.5 mm implants were most stable with a mean periotest value of −1.80 ± 0.0 while 10 mm implants had the highest mean periotest score of 1.72 ± 9.57. The reason for this may be due to the low number of 11.5 mm implants used in this study. A negative correlation between implant length and implant stability was observed, but this was not statistically significant. Our finding is in contrast with a report which revealed a positive significant correlation between implant length and implant stability.[10] The reason for the variation in our findings may be due to the differences in the location of the implants in both studies. Their implants were placed mainly in the posterior mandible alone which is the best anatomic site to obtain maximum implant stability,[10] while in the present study, the implants were placed in all segments of the jaws.

We further investigated the relationship between implant diameter and implant stability. It was observed that 4.7 mm implants had the lowest mean periotest value (−0.67 ± 4.20) followed by 4.1 mm implants (0.76 ± 7.42) and 3.7 mm implants (2.13 ± 1.55). There was a negative correlation between implant diameter and implant stability which was not statistically significant and is in agreement with previous studies.[10],[11] Previous studies have documented that implant diameters are positively correlated to implant stability. This is because of the larger contacting surface of the implant with bone. This study achieved a high primary implant stability for all cases (mean torque = 45 N-cm) as a result of the careful osteotomy preparation as well as the use of tapered and screw implant designs. It has been reported that screw design improves primary stability, the principal requirement for immediate loading success.[12],[13],[14],[15]

There was a statistically significant association between initial periotest value at placement and implant survival, indicating that the periotest value at placement affects the 1-year survival rate of the implants. This is in agreement with what is obtained in other reports.[16],[17] An initial periotest value is necessary for implant survival as it shows that the implant is well integrated into the bone with little or no mobility and can be loaded deterring prosthetic loading of an unstable implant.

Significantly, lower mean periotest value was recorded in the mandible (−2.12) compared to the maxilla (1.95), indicating better implant stability in the mandible than in the maxilla. This is in agreement with the findings of majority of similar studies.[18],[19] Our result is however in contrast with the findings of a study carried out by Oh et al.,[20] which showed better periotest values in the maxilla than in the mandible. The reason for this may be attributed to the different study subjects as animal subjects (mongrel dogs) used by the investigators as opposed to human subjects used in our study.[20] Further, the better (low) periotest value obtained in the mandible may be due to the better volume and density of bone available for osseointegration.[19]

In the maxilla, lower periotest values were obtained in the posterior region compared to the anterior region although this was not statistically significant. Similarly, in the mandible, lower but insignificant periotest values were obtained in the posterior region compared to anterior region. These results are in contrast with a study by Cranin et al.,[10] where lower mean periotest values were obtained in the anterior segments of both the maxilla and mandible. The reason for this may be attributed to a significantly large number of implants placed in the anterior region of the maxilla and mandible in the present study as opposed to the posterior region.


   Conclusion Top


There is a definite relationship between implant characteristics and implant stability of immediately loaded implants. This is denoted by a negative correlation between implant length, diameter, and the periotest values, denoting that implant stability increases as the length and diameter increases. Immediately loaded implants placed in the mandible tend to be more stable than those placed in the maxilla, with the initial stability at implant placement having a significant association with the implant stability after 1-year of follow-up.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Schnitman PA, Wohrle PS, Rubenstein JE. Immediate fixed interim prostheses supported by two-stage threaded implants: Methodology and results. J Oral Implantol 1990;16:96-105.  Back to cited text no. 1
    
2.
Testori T, Del Fabbro M, Galli F, Francetti L, Taschieri S, Weinstein R, et al. Immediate occlusal loading the same day or the after implant placement: Comparison of 2 different time frames in total edentulous lower jaws. J Oral Implantol 2004;30:307-13.  Back to cited text no. 2
    
3.
Testori T, Meltzer A, Del Fabbro M, Zuffetti F, Troiano M, Francetti L, et al. Immediate occlusal loading of osseotite implants in the lower edentulous jaw. A multicenter prospective study. Clin Oral Implants Res 2004;15:278-84.  Back to cited text no. 3
    
4.
Christopher CK. Immediate function with dental implants. Oral Implants 2005;158:7.  Back to cited text no. 4
    
5.
Roberts WE, Smith RK, Zilberman Y, Mozsary PG, Smith RS. Osseous adaptation to continuous loading of rigid endosseous implants. Am J Orthod 1984;86:95-111.  Back to cited text no. 5
    
6.
Szmukler-Moncler S, Piattelli A, Favero GA, Dubruille JH. Considerations preliminary to the application of early and immediate loading protocols in dental implantology. Clin Oral Implants Res 2000;11:12-25.  Back to cited text no. 6
    
7.
Søballe K. Hydroxyapatite ceramic coating for bone implant fixation. Mechanical and histological studies in dogs. Acta Orthop Scand Suppl 1993;255:1-58.  Back to cited text no. 7
    
8.
Vaillancourt H, Pilliar RM, McCammond D. Finite element analysis of crestal bone loss around porous-coated dental implants. J Appl Biomater 1995;6:267-82.  Back to cited text no. 8
    
9.
Medizintechnik Gulden. The Periotest; How Does it Work. Available from: https://www.med-gulden.com/periotest.php?t=lnk_2#lnk_2. [Last assessed on 2016 Dec 11].  Back to cited text no. 9
    
10.
Cranin AN, DeGrado J, Kaufman M, Baraoidan M, DiGregorio R, Batgitis G, et al. Evaluation of the periotest as a diagnostic tool for dental implants. J Oral Implantol 1998;24:139-46.  Back to cited text no. 10
    
11.
Lorenzoni M, Pertl C, Zhang K, Wimmer G, Wegscheider WA. Immediate loading of single-tooth implants in the anterior maxilla. Preliminary results after one year. Clin Oral Implants Res 2003;14:180-7.  Back to cited text no. 11
    
12.
Hansson HA, Albrektsson T, Brånemark PI. Structural aspects of the interface between tissue and titanium implants. J Prosthet Dent 1983;50:108-13.  Back to cited text no. 12
    
13.
Skalak R. Aspects of biomechanical considerations. In: Branemark PI, Zarb G. Tissue-Integrated Prosthesis: Osseointegration in Clinical Dentistry. Quintessence Books:Michigan, USA; 1985. p. 117-28.  Back to cited text no. 13
    
14.
Wolfe LA, Hobkirk JA. Bone response to a matched modulus endosseous implant material. Int J Oral Maxillofac Implants 1989;4:311-20.  Back to cited text no. 14
    
15.
Misch CE. Implant design considerations for the posterior regions of the mouth. Implant Dent 1999;8:376-86.  Back to cited text no. 15
    
16.
Negm SA. Implant success versus implant survival. J Dent 2016;6:359.  Back to cited text no. 16
    
17.
Van Scotter DE, Wilson CJ. The periotest method for determining implant success. J Oral Implantol 1991;17:410-3.  Back to cited text no. 17
    
18.
Olivé J, Aparicio C. Periotest method as a measure of osseointegrated oral implant stability. Int J Oral Maxillofac Implants 1990;5:390-400.  Back to cited text no. 18
    
19.
Turkyilmaz I, McGlumphy EA. Influence of bone density on implant stability parameters and implant success: A retrospective clinical study. BMC Oral Health 2008;8:32.  Back to cited text no. 19
    
20.
Oh JS, Kim SG, Lim SC, Ong JL. A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;107:513-8.  Back to cited text no. 20
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7], [Table 8], [Table 9], [Table 10], [Table 11]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed4830    
    Printed97    
    Emailed0    
    PDF Downloaded227    
    Comments [Add]    

Recommend this journal